Test Name BF Alb Albumin Level Body Fluid
Specimen Requirements
Preparation of Patient:
None
Container Type:
Sterile Container
Specimen Type:
Peritoneal Fluid
Note: Indicate specimen source
Specimen Volume:
3 mL of body fluid
Specimen Handling/Transport:
Fluid should be removed from the cells within 1 hour of collection.
Transport refrigerated 2 - 8oC.
Specimen Stability/Storage:
5 days refrigerated 2-8°C
Specimen Rejection:
pH outside 7-8 range
Performing Laboratory
Munson Medical Center Laboratories
Priority, Frequency, & Turnaround
Priority:
STAT, ASAP, Non-Emergent
Frequency:
24 hours 7 days a week
Turnaround:
0 - 1 days
STAT 60 minute from receipt in lab
Methodology
Bromcresol Green
Reporting
Reference Range:
A serum to effusion albumin gradient (serum albumin minus fluid albumin) can be helpful in determination of a transudate or exudate. A values less than 1.1 g/dL is seen in exudates. This is a laboratory-developed test.
Linearity Range:
1.5 - 15.0 g/dL
Critical Decision:
None
Clinical Significance
A serum to effusion albumin gradient (serum albumin minus fluid Albumin) can be helpful in determination of a transudate or exudate. A value less than 1.1 g/dL is seen in exudates.
CPT Code(s)
82042